AMV

Analytical Method validation of Drench

QA SOPS

1.0 Method Validation:

Analytical method validation is the process to confirm that the analytical procedure employed for a specific test is suitable for its intended use.

2.0 Objective:

Analytical monitoring of a pharmaceutical product is necessary to ensure its efficacy throughout all phases of its shelf life; such monitoring is in accordance with the specifications elaborated during product development. Method validation ensures that the selective method will give reproducible, reliable, and consistent results adequate for the intended purpose, it is, therefore, necessary to define precisely both the conditions in which the procedure is to be used and the purpose for which it is intended.

The objective of this validation study is to;

  • Ensure and justify, through extensive testing, that the precision and consistency are in accordance with already established Acceptance Criteria.
  • Assess the effect of variables (within already set operating limits) on the testing method
  • Identify and solve the problem(s), if any, encountered during testing.
  • Establish confidence in the existing testing process.
  • Find ways and means to increase productivity and improve quality.
  • Assure that the complete process is under control.

3.0 Scope:

The Scope of this document is

  1. To describe the work required for the Validation of the Test Method for the testing of Lumenil Drench, This method is routinely used at the time of manufacturing and during stability studies.
  2. This Validation Protocol also describes the analytical parameters to be used for the validation of the test method.
  3. This Protocol also defines the facilities, responsibilities and equipment, apparatus, glassware, material, and documents that are used for the validation studies.
  4. In this validation protocol, the analytical results are evaluated by the application of statistical techniques and presented by means of graphical techniques.
  5. This method validation protocol applies to all test methods performed for release or stability evaluation of Lumenil Drench.

4.0  Periodic Revalidation:

In case there is any change or modification in the Validated Method, the Revalidation will be performed.

5.0  Change Control:

Any modification or changes in the Validated Test Method (SOP No. /IMS/SAP/QC/018)being used for the testing of  Lumenil  Drench should be controlled and will be entered into the Change Control Form (Form NoIR/QC/0–) in accordance with the change control procedure (SOP No /IMS/SAP/QC/0-)

Provide Justification for proposing a test method. The proposed procedure should also be validated according to this Validation Protocol.

Provide Justification for proposing a test method. The proposed procedure should also be validated according to this Validation Protocol.

 

6.0  Facilities/Responsible Personnel/Equipment/Material Documentation

 

6.1  Facilities:

Validation of Test Method (SOP No. /IMS/SAP/FP/00) being used for the Determination of Oxfendazole & Oxyclozanide in Lumenil  Drench will be performed in Chemical Lab.

 

  • Machine/ Equipment
Machine/ Equipment Qualification Status Qualification No. Calibration Status
HPLC Qualified IQ:     _________

OQ:    _________

PQ:     _________

 

 

 

Analytical  Balance Qualified IQ:     _________

OQ:    _________

PQ:     _________

 

 

 

 

  • Identification of Apparatus/Glassware
Apparatus/Glassware
Volumetric Flask
Pipette

 

7.0 Analytical Method Of Lumenil  Drench

As Per FP SOP NO /IMS/SAP/FP/0

Physical Parameters:

 

PARAMETERS SPECIFICATION
Physical Form Suspension
Colour Cream Colored
.pH 3.0 to 6.0 at 25° C
Viscosity 1200 – 1500 C.P.S  at 25°C

Spindle # LV 2, 50 rpm,Tarq NLT 40%

Wt/ml 1.04gm/ml

 

Chemical Analysis

PARAMETERS SPECIFICATION
Identification Test of Oxfendazole

By retention time of std and sample  in the chromatograph

The retention time of std and sample in chromatograph should be the same.
Identification Test of Oxyclozanide

By retention time of std and sample  in the chromatograph

The retention time of std and sample in chromatograph should be the same.

Chemical Analysis

  1. Contents of Oxyclozanide:

Stated Quantity: 62.50 mg / ml

  1. Contents of Oxfendazole:

Stated Quantity:  22.50 mg / ml

1.1 Sample Preparation:

NOTE: Before taking sample shake well and sonicate suspension for 10minutes in a 250ml glass beaker.

Take 5.20gm of a sample of suspension, previously well mixed, sonicated(10minutes), and free from air bubbles equivalent to 112.5 mg of Oxfendazole and 312.5mg oxyclozanide in 100ml glass beaker, add 20ml distilled water in a beaker in which sample is taken and sonicate it for 10 minutes, then transfer this sample completely in 100ml volumetric flask. rinse the same beaker with a small portion of methanol and transfer to a flask in which the sample is taken.

Add methanol in portions while sonicating and shaking the flask,and makeup  volume up to 100ml with methanol.

  1. Shake well and Sonicate 100ml volumetric flask (step a) for 30minutes till clear solution and filter it.
  2. c) Take 1 ml from (step b) filtrate in 50 ml volumetric flask and make volume upto 50 ml with methanol.mix well and sonnicate for 5minutes.

1.2  Oxfendazole Standard Preparation:

 

  1. Weight accurately 113 mg Oxfendazole and put them in a 100 ml Volumetric flask. Then add 20 to 30 ml methanol ,sonnicate flask for complete solubility  and make volume upto 100 ml with methanol.
  2. Shake well and Sonicate 100ml volumetric flask (step a) for 10 minutes.
  3. Take 1 ml from (step b) in 50 ml volumetric flask and make volume upto 50 ml with methanol and sonnicate for 5minutes

1.3 Oxyclozanide Standard Preparation:

  1. Weight accurately 312 mg oxyclozanide and put them in a 100 ml Volumetric flask. Then add 20 to 30 ml methanol.sonnicate  flask for complete solubility  and make volume upto 100 ml .
  2. Shake well and Sonicate 100ml volumetric flask (step a) for 10 minutes .
  3. Take 1 ml from (step b) in 50 ml volumetric flask and make volume upto 50 ml with methanol and sonnicate for 5minutes.

1.3 Chromatographic system

The liquid chromatograph is equipped with octadecyl silane column of C18  250mm X            4.6mm .

Flow Rate:                            1 ml / min.

Detector:                              UV 300 nm.

1.4 Mobile Phase:

1% H3PO4 (OrthoPhosphoric Acid)   :  Methanol (HPLC grade)

20 ml                                            :    80 ml

Note : 1% H3PO4 (Phosphoric Acid):  [ 1ml Con phosphoric acid + Sufficient Distilled water to make total volume 100ml]

1.5 Calculations:

  1. Contents of Oxyclozanide:

Stated Quantity: 62.50 mg / ml

 

Calculations:

Ave. Peak area  of Sample =__________________________

Ave. Peak area  of Standard =___________________________

 

Ave. Peak area Sample 312   X  X 100  X 50 X 1.04(wt/ml) =–X——– mg/ml
Ave. Peak area Standard      100 50  5.2 gm  1
    X      X 100 =———%
  62.5

 

Limit: The contents of Oxyclozanide should be between 90 -110% of stated amounts.

2.0 Contents of Oxfendazole:

Stated Quantity:  22.50 mg / ml

Procedure: Inject standard 1st.

  1. Make 2 consecutive injections of standard preparation.
  2. Make 2 consecutive injections of sample preparation.

Calculations:

Ave. Peak area  of Sample =__________________________

Ave. Peak area  of Standard =___________________________

Ave. Peak area Sample 113   X  X 100  X 50 x  1.04(wt/ml =–X——– mg/ml
Ave. Peak area Standard      100 50 5.2gm  1
    X      X 100 =———%
  22.50

 

Limit: The contents of Oxfendazole should be between 90 -110% of stated amounts

 

  • Validation Parameters: Following analytical parameters are to be considered
  • Linearity:(Calibration Curve )
  • Precision
    • System precision (Repeatability)
    • Method precision (Reproducibility)
    • Intermediate Precision:
  • Accuracy
  • Ruggedness
  • Robustness:
  • Stress Degradation:
  • Limit of Detection (LOD):
  • Limit of Quantitation (LOQ):
  • Specificity
  • system Suitability:

8.1 Calibration Curve (Linearity):

Linearity is the ability of the method to produce results that are directly proportional to the concentration of the analyte in samples with a given range.

8.1.1 Calibration Curve (Linearity) of Oxfendazole

The linearity of the method was evaluated by preparing three series of standard of Oxfendazole in the range of  70 – 130 μg/ml in methanol and note peak area  from chromatograph .Excellent correlation between peak area and concentration was observed with R2 = 0.9971 (Figure.1). The regression equation was found to be Y=315706x+472085.Statistical data are presented in table 1and the calibration curve was shown in figure 1.

Table 1: Linearity results for Oxfendazole

 

Sr. # Chrom  # Conc. (ug/ml) Peak area  MEAN RSD (%)
1 AMV 01 70 775424 777985.0

 

 

0.710%
2 AMV 02 774210
3 AMV 03 784321
4 AMV 01 100 1121530 1123110.3

 

 

0.122%

5 AMV 02 1123861
6 AMV 03 1123940
7 AMV 01 130 1411486 1409397.7

 

0.147%
8 AMV 02 1409355
9 AMV 03 1407352

 

Figure 1: Calibration curve for Oxfendazole

Acceptance Criteria:

  • Coefficient of determination (r2) should be greater than 0.9998.

8.1.1 Calibration Curve (Linearity) of Oxyclozanide

 

The linearity of the method was evaluated by preparing three series of standard  of Oxyclozanide in the range of  70 – 130 μg/ml in methanol and note peak area  from chromatograph .Excellent correlation between peak area and concentration was observed with R2 = 0.9995 (Figure.1). The regression equation was found to be Y=581220x+ 704039 Statistical data are presented in table 1and the calibration curve was shown in figure 1.

Table 2: Linearity results for Oxyclozanide

Sr. # Chrom  # Conc. (ug/ml) Peak area  MEAN RSD (%)
1 AMV 01 70 1293250 1277493.0

 

 

1.152%
2 AMV 02 1264111
3 AMV 03 1275118
4 AMV 01 100 1899469 1882009.7

 

 

1.169%

5 AMV 02 1889269
6 AMV 03 1857291
7 AMV 01 130 2449197 2439932.7

 

2.255%
8 AMV 02 2489739
9 AMV 03 2380862

 

Figure 1: Calibration curve for Oxyclozanide

Acceptance Criteria:

  • Coefficient of determination (r2) should be greater than 0.9998.
  • Precision:

Precision is the degree of closeness of agreement among individual test results when the method is applied to multiple sampling of a homogeneous sample. It is a measure of either the degree of reproducibility (under different conditions) or repeatability (under same conditions) of the method.

8.2.1 System precision (Repeatability)

To study precision, five replicate standard of Oxfendazole (20μg/ml) were prepared and analyzed using the proposed method. The percent relative standard deviation (% RSD) for peak responses was calculated. Results of system precision studies were shown in table 3.

Results of system precision for Oxfendazole, Absorbance repeatability

Table 3
Sr. # # Peak area
001 AMV Repeat 01 1113443
002 AMV Repeat 02 1106940
003 AMV Repeat 03 1107423
004 AMV Repeat 04 1113836
005 AMV Repeat 05 1113300
n = 5
Mean 1110988.40
SD 3484.93
RSD % 0.314%

 

8.2.1 System precision (Repeatability)

To study precision, five replicate standard of Oxyclozanide (20μg/ml) were prepared and analyzed using the proposed method. The percent relative standard deviation (% RSD) for peak responses was calculated. Results of system precision studies were shown in table 4.

Results of system precision for Oxyclozanide, Absorbance repeatability

Table 4
Sr. # # Peak area
001 AMV Repeat 01 2008905
002 AMV Repeat 02 1973729
003 AMV Repeat 03 1972313
004 AMV Repeat 04 1963391
005 AMV Repeat 05 2002173
n = 5
Mean 1984102.20
SD 20107.68
RSD % 1.013%

 

8.2.2 Method precision (Reproducibility)

The intraday and inter-day precision of the proposed method was determined by analyzing the corresponding responses 2 times on the same day and on different days for concentration of standard of 20μg/ml. The result was reported in terms of relative standard deviation (% RSD). Results of method precision studies were shown in table 5.

Results of Results of Method precision for Oxfendazole

                                                       Table 5

  Standard preparation 1 Standard preparation 2
Sr. # Chrom. # Peak Area Chrom. # Peak Area
001 1115941 1124759
002 1119606 1125140
n =   2   2
Mean   1117773.50   1124949.50
RSD %   0.232%   0.024%
 
Grand Mean 1121361.50  
RSD % 0.453%  

 

There were no significant differences between RSD (%) values for intra-day and inter-day precision, which indicates that the method is reproducible.

8.2.3 Method precision (Reproducibility)

 

The intraday and inter-day precision of the proposed method was determined by analyzing the corresponding responses 2 times on the same day and on different days for concentration of standard  of 20μg/ml. The result was reported in terms of relative standard deviation (% RSD). Results of method precision studies were shown in table 6.

Results of Results of Method precision for Oxyclozanide  

                                                       Table 6

  Standard preparation 1 Standard preparation 2
Sr. # Chrom. # Peak Area Chrom. # Peak Area
001 1918711 1930483
002 1929929 1930649
n =   2   2
Mean   1924320   1930566
RSD %   0.412%   0.006%
 
Grand Mean 1927443  
RSD % 0.229%  

 

There were no significant differences between RSD (%) values for intra-day and inter-day precision, which indicates that the method is reproducible.

8.2.3  Intermediate precision

The intermediate precision of the proposed method was determined by performing the method by two analysts (Analyst 1 and Analyst 2) for concentration of standard 20μg/ml. The percent relative standard deviation (% RSD) for peak responses was calculated. The results for intermediate precision were shown in table 7,8.

Intermediate precision refers to variations within a laboratory as with, different instruments, by different analysts, and so forth.

 

Results of Intermediate precision for Oxfendazole

Table 7
ANALYST 1
Instrument: HPLC System 01
Performed By          (QCM)
Chrom  #  
1127518
1130361
n 2
Mean 1128939.500
SD 2010.3046
RSD % 0.178%

 

Table 8
ANALYST 2
Instrument: HPLC System 02
Performed By  (QC analyst)
Abs #  
1130742.0
1137243.0
n = 2  
Mean 1133992.500
SD 4596.9012
RSD % 0.405%

 

8.2.4  Intermediate precision

The intermediate precision of the proposed method was determined by performing the method by two analysts (Analyst 1 and Analyst 2) for concentration of standard 20μg/ml. The percent relative standard deviation (% RSD) for peak responses was calculated. The results for intermediate precision were shown in table 9,10.

Intermediate precision refers to variations within a laboratory as with, different instruments, by different analysts, and so forth.

Results of Intermediate precision for Oxyclozanide

Table 9
ANALYST 1
Instrument: HPLC System 01
Performed By          (QCM)
Chrom  #  
1954688
1917003
n 2
Mean 1935845.50
SD 26647.32
RSD % 1.377%

 

Table 10
ANALYST 2
Instrument: HPLC System 02
Abs #  
1898317.0
1939108.0
n = 2  
Mean 1918712.50
SD 28843.5927
RSD % 1.503%

 

  • Accuracy ( % Recovery):

 

Accuracy is the closeness of results obtained by a method to the true value. It is the measure of exactness of the method. The accuracy of the method was evaluated by the standard addition method. Recovery of the method was determined by spiking an amount of the pure drug (50%,100%,150%) at three different concentration levels in its Drench has been added to the pre-analyzed working standard Drench of the drug. The sample was analyzed in triplicate at each level. The percent individual recovery and %RSD for recovery at each level are calculated.

  • Accuracy ( % Recovery) of Oxfendazole

Peak area of Oxfendazole RS

1128939.500

 

 

Sample preparations

 

Table 11
 
Level of recovery Initial amount of SA (mg) Amount of std added
(mg)
Total amount present (mg) Chrom # Peak area Mean area Total amount recovered (mg) Recovery

(%)

50% 20 30 50 459073 457846.000

 

50.69 101.39%
456619
100% 60 40 100 621691 623372.000

 

99.39 99.39%
625053
150% 100 50 150 790120 787539.500

 

149.98 99.99%
784959
Mean Recovery   100.26%
SD 1.025
RSD % 1.022%

 

Acceptance Criteria

Recovery                      100 + 2%

The recovery of the method is determined by spiking a placebo with active pharmaceutical ingredient of known potency. The values of recovery (%), RSD (%),in Table 11, above indicate that the method is accurate.

  • Accuracy ( % Recovery) of Oxyclozanide

Peak area of Oxyclozanide RS

1935845.500

 

 

Sample preparations

 

Table 12
 
Level of recovery Initial amount of SA (mg) Amount of std added
(mg)
Total amount present (mg) Chrom # Peak area Mean area Total amount recovered (mg) Recovery

(%)

50% 20 30 50 783612 776757.000

 

50.16 100.13%
769902
100% 60 40 100 1045308 1041968.500

 

99.58 99.58%
1038629
150% 100 50 150 1321999 1326721.000

 

148.72 99.15%
1331443
Mean Recovery   99.68%
SD 0.590
RSD % 0.592%

 

Acceptance Criteria

Recovery                      100 + 2%

The recovery of the method is determined by spiking a placebo with active pharmaceutical ingredient of known potency. The values of recovery (%), RSD (%),in Table 12, above indicate that the method is accurate.

 

  • Ruggedness:

 

The ruggedness of an analytical method is the degree of reproducibility of test results obtained by the analysis of the same samples under a variety of conditions, such as different laboratories, different analysts, different instruments, different lots of reagents, different elapsed assay times, different assay temperatures, different days, etc.  Ruggedness is normally expressed as the lack of influence on the test results of operational and environmental variables of the analytical method.  Ruggedness is a measure of reproducibility of test results under the variation in conditions normally expected from laboratory to laboratory & from analyst to analyst.

Procedure:

  • Analyze two samples from the same composite sample according to the analytical method by two analyst using different lots of reagents keeping the other condition same.
  • Analyze the samples according to procedure by one analysts on the same day & the other on second day.

For Oxfendazole

Table 13

  

                      Day 1                                       

ANALYST 1
For Oxfendazole

Standard =

1128939.500

Sr Peak area Theoretical
Content
Amount recovered Recovery (%)
1 5246057.000 22.50mg/ml 22.37mg/ml 99.43%
2 5277187.000
Mean 5261622.000
SD 22012.2341
%RSD 0.418%
Acceptance Criteria:

 

The % RSD of the assay or recovery values should not be greater than 2.0%.
Assay Limit 97 % to 103 %

 

Table 14

  

                      Day 2                                           

 

ANALYST 2

For Oxfendazole

Standard =

1133992.500

Sr Peak area Theoretical
Content
Amount recovered Recovery (%)
1 5100169.000 22.50mg/ml 22.65mg/ml 100.68%
2 5257771.000
Mean 5178970.000
SD 111441.4429
%RSD 2.1518
Acceptance Criteria:

 

The % RSD of the assay or recovery values should not be greater than 2.0%.
Assay Limit 97 % to 103 %

 

For Oxyclozanide

Table 15

  

                      Day 1                                           

 

ANALYST 1

 

For Oxyclozanide

Standard =

1935845.500

Sr Peak area Theoretical
Content
Amount recovered Recovery (%)
1 10125601.000 62.50mg/ml 63.08mg/ml 100.93%
2 10227873.000
Mean 10176737.000
SD 72317.2247
%RSD 0.711%
Acceptance Criteria:

 

The % RSD of the assay or recovery values should not be greater than 2.0%.
Assay Limit 97 % to 103 %

 

Table 16

  

                      Day 2                                           

 

ANALYST 2

For Oxyclozanide

Standard =

1918712.500

Sr Peak area Theoretical
Content
Amount recovered Recovery (%)
1 9934531.000 62.50mg/ml 62.92mg/ml 100.67%
2 9856607.000
Mean 9895569.000
SD 55100.5888
%RSD 0.5568%
Acceptance Criteria:

 

The % RSD of the assay or recovery values should not be greater than 2.0%.
 

Assay Limit 97 % to 103 %

 

The ruggedness of the proposed method was determined by two different analysts. The mean, standard deviation, and % RSD values were found to be less than 2 % which shows the ruggedness of the method.

  • Robustness:

The robustness study was performed to evaluate the influence of small but deliberate variation in the chromatographic condition. The robustness was checked by changing parameters like flow rate of mobile phase and detection wavelength  Change in the detection wavelength by ± 2nm (298nm and 302nm)  Change in flow rate by ± 0.1 ml/minute (0.6 ml/min and 0.8 ml/minute) After each change, sample was injected and % assay with system suitability parameters were checked. Robustness values were given in table 17&18.

 

Results of Robustness values for Oxfendazole

Table17

  

 Changing Parameters

 

Detection Wavelength

Instrument HPLC SYSTEM 01
Model /Made WATERS
For Peak area

Standard =

1128939.500

Sr Wavelength Peak area Theoretical
Content
Amount recovered Recovery (%)
1 298 5471089 22.50mg/ml 22.89mg/ml 101.73%
2 300 5396254
3 302 5417940
Mean 5428427.66
SD 38504.06
%RSD 0.709%
Acceptance Criteria:

 

The % RSD of the assay or recovery values should not be greater than 2.0%.
Assay Limit 97 % to 103 %

 

Results of Robustness values for Oxyclozanide

Table18

  

 Changing Parameters

 

Detection Wavelength

Instrument HPLC SYSTEM 01
Model /Made WATERS
For Peak area

Standard =

1918712.500

Sr Wavelength Peak area Theoretical
Content
Amount recovered Recovery (%)
1 298 10016338 62.50mg/ml 63.51mg/ml 101.61%
2 300 9970189
3 302 9977307
Mean 9987944.667
SD 24845.5733
%RSD 0.249%
Acceptance Criteria:

 

The % RSD of the assay or recovery values should not be greater than 2.0%.
Assay Limit 97 % to 103 %

 

Robustness studies were made by varying the chromatographic parameters like detector wavelength and flow rate are within a realistic range and the quantitative influence of the variables is determined. The robustness of the proposed method was determined by estimating a drug at a slightly different wavelength from the selected wavelength and a slight change in flow rate. No significant difference was found in the RT & parentage of samples. Therefore, the proposed method was considered robust.

  • Stress Degradation:

Degradation information obtained from stress studies (e.g., products of acid and base hydrolysis, thermal degradation, photolysis and oxidation) for the drug substance and for the active ingredient in the drug product.

Procedure (Thermal Degradation):

  • Prepare samples Drench from the same assay composite sample according to the analytical method by one analyst at room temperature
  • Prepare samples Drench, and standard Drench according to the analytical method by one analyst and subjected to heat treatment at, 60 °C for one hour
  • Analyze both the samples.

For Oxfendazole

Table 19

    UNSTRESS SAMPLE
For Oxfendazole

Standard =

1128939.500

Sr Absorbance Theoretical
Content
Amount recovered Recovery (%)
1 5281553.000 22.50mg/ml 22.502mg/ml 100.01%
2 5303048.000
Mean 5292300.500
SD 15199.2603
%RSD 0.287%
Acceptance Criteria:

 

The % RSD of the assay or recovery values should not be greater than 2.0%.
Assay Limit 97 % to 103 %

 

Table 20

   FORCED DEGRADATION
For Oxfendazole

Standard =

1128939.500

Sr Absorbance Theoretical
Content
Amount recovered Recovery (%)
1 5259868 22.50mg/ml 22.36mg/ml 99.39%
2 5259868
Mean 5259868
SD 0.0000
%RSD 0.000
Acceptance Criteria:

 

The % RSD of the assay or recovery values should not be greater than 2.0%.
Assay Limit 97 % to 103 %

 

For Oxyclozanide

Table 21

   UNSTRESS SAMPLE
For Oxyclozanide

Standard =

1935845.500

Sr Absorbance Theoretical
Content
Amount recovered Recovery (%)
1 9961648.000 62.50mg/ml 62.90mg/ml 100.64%
2 9999654.000
Mean 9980651.000
SD 26874.3003
%RSD 0.269%
Acceptance Criteria:

 

The % RSD of the assay or recovery values should not be greater than 2.0%.
Assay Limit 97 % to 103 %

 

Table 22

   FORCED DEGRADATION
For Oxyclozanide

Standard =

1935845.500

Sr Absorbance Theoretical
Content
Amount recovered Recovery (%)
1 9815505 62.50mg/ml 62.62mg/ml 100.19%
2 10056584
Mean 9936044.500
SD 170468.5957
%RSD 1.716
Acceptance Criteria:

 

The % RSD of the assay or recovery values should not be greater than 2.0%.
Assay Limit 97 % to 103 %

 

There was no degradation of Oxfendazole & Oxyclozanide and no significant change in % of Oxfendazole & Oxyclozanide.

  • Limit of detection (LOD) & Limit of Quantitation (LOQ)

The LOD and LOQ were predicted based on the parameters of the standard error of estimate and slope, calculated from the linearity of the response data.

  • Limit of Detection (LOD):

The lowest amount of analyte in the sample can be detected but not necessarily quantitated. Analysis of sample with the known concentration of analyte and by establishing minimum concentration at which analyte can be reliably detected.

Procedure:

LOD may also be calculated based on the standard deviation (SD) of the response and the slope of the calibration curve (S) at levels approximating the LOD according to the formula: LOD =3.3(SD/S). The standard deviation of the response can be determined based on the standard deviation of y-intercepts of regression lines.

In formula;

S = Slop of Calibration Curve

SD = Standard Deviation of response

 

LOD          =              3.3     X       SD

Slope 

 

Limit of Quantitation (LOQ):

The lowest amount of an analyte in a sample can be quantitatively determined with suitable precision and accuracy. Analysis of sample with the known concentration of analyte and by establishing minimum concentration at which analyte can be reliably quantitated.

LOQ may also be calculated based on the standard deviation of the response (SD) and the slope of the calibration curve (S) at levels approximating the LOQ according to the formula: LOQ =10(SD/S). The standard deviation of the response can be determined based on the standard deviation of y-intercepts of regression lines.

In formula;

S = Slop of Calibration Curve

SD = Standard Deviation of response

 

LOQ         =              10      X     SD

Slope

 

LIMIT OF DETECTION (LOD) AND LIMIT OF QUANTITATION (LOQ) OF Oxfendazole & Oxyclozanide

Table 23

 

SR # Compound SLOPE STEXY

 

       LOD        LOQ
(known_y”s,known_X”s,)  (known_y”s,known_X”s,) = 3.3 X  SD

Slope 

= 10 X  SD

Slope 

1 Oxfendazole

 

315706.350 24020.472 0.228 0.761
2 Oxyclozanide 581219.850 19021.798 0.098 0.327

 

 Specificity

The specificity of an analytical method is the ability to assess unequivocally the analyte in the presence of components that may be expected to be present, such as impurities, degradation products, and matrix components.

  • Poor specificity can impact accuracy, precision, and linearity

Acceptance Criteria:

  • Assay Limit 98 % to 102 %
  • The coefficient of determination (r2) should be greater than 0.9998.
  • The % RSD of the assay or recovery values should not be greater than 2.0%.

Remarks: Accuracy, precision, and linearity are within the limit, it confirms that the analytical procedure employed for a specific test is suitable for its intended use.

  System Suitability:

Set of parameters and criteria there off to ensure that the system is working properly. System suitability conforms to repeatability data. (Mean, Standard Deviation, and Relative Standard Deviation (%).

The performance of the instrument is satisfactory because all the parameters of the system suitability test (Rt, ReDrench, Theoretical Plates,) are within the acceptance range.

 

CONCLUSION

The developed method was found to be sensitive, accurate, precise, reproducible, and linear over the concentration range studied. The proposed method can be used for the routine quality control

analysis of Lumenil  Drench.

 

Sign:

Date:

Q.C Analyst

 

Sign:

Date:

Quality Control Manager

 

if you have any comment/suggestion about this article ”Analytical Method validation of Drench” then do write to us.